Abstract
AimLoosing of cerclage wire is a common problem in the fixation of fractures and alternative techniques are required to provide more strength cable systems. The aim of this study was to evaluate the benefit of an additional knot to the cable system to increase durability and strength of wireMethodsTwelve artificial bone fragments were bisected thoroughly in the coronal plane. Two mm diameter CoCr Dall Miles were used in two groups; the first group was the sleeve group (n = 6) and the second group was the sleeve + knot group (n = 6). All cables were once looped around the saw bone with positioning the sleeve at the beaded end of the cable, and the tensioner knob was used to create tension to 90 lb/in2 (62,05 N/cm2). In the sleeve + knot group, one more simple knot was performed with conventional surgery skill, and the two ends of cerclage wire were put into the tensioner knob to create tension again. Mechanical tension testing was performed at a velocity of 20mm/min. Axial tension force was applied until failure. The maximum force and displacement values were recorded at the failure point.ResultsThere was a significant difference in the mean rank of applied force between the sleeve and sleeve + knot groups (P < .001,025). However, there was no statistical difference in the amount of displacement between the two groups (P < .001,378). The ratio of force/displacement displayed a significant difference between the two groups, it was shown that the sleeve + knot group was more durable than the sleeve only group (P < .001,01). ConclusionAdding a knot to the cable system increases the durability and strength of the cable system by approximately %40. These findings showed that adding a knot to the cable-domino system had decreased the possibility of loosening and fixation failure.
Keywords
Introduction
Cerclage wiring, a basic surgical technique, has been used in many areas of orthopedic surgeries, especially to help or support the fixation of fractures mainly long bones and periprosthetic fractures. In the past, many materials such as titanium alloy, chrome-cobalt, stainless steel were used for cable systems.1
Documentation of the first use of cerclage wire dates back to 1775.2 It has become an option in different areas such as fixation of unstable intertrochanteric, acetabular, periprosthetic fractures of femur, humerus, elbow, revision total hip arthroplasties in combination with other implants.3-9 In recent years, the use of cerclage wire has increased as a result of the increase of periprosthetic fractures, and in this situation there is increasing interest in the cerclage wire techniques, which provide augmentation and make fixation more stable. Surgeons try to overcome the insufficiency and stability of cerclage by using different techniques and adding other implants such as Steinman pins,10 external plaster cast.9 Since simple cerclage wiring alone is related to high failure rates, fixing the rigidity is very important.11 Therefore, researchers are focusing on the factors that affect yield and breaking load, fatigue strength because the best way to measure mechanical strength is to evaluate yield and breaking loads. Fatigue strength is also useful for the measurement of the strength of cerclage wire.1
The multifilament cable was first used for the fixation of the greater trochanter osteotomy in total hip arthroplasty by Dall and Miles.12The widely used Dall Miles cables are made of stainless steel multifilament and are twice as strong as surgical chrome wire.10,13 . Other beneficial effects of multifilament cables are preventing kinking, which is widely accepted as a risk factor for breakage of cable, making control of tension degree easier.6,13,14 However, surgeons also encounter failure of multifilament cables because of the patient’s biologic factors (bone quality, comorbidities, life qualities), the surgeons’ insufficiency about instruments stability, and also wire itself can fail.
In literature, the best cerclage wire configuration is still unclear. This study compared the failure of a load of conventional and our augmentation technique by detecting the strength point value when displacement had occurred.
Materials and Methods
Study GroupsNo animal or human studies were carried for this study so there was no need for ethical approval. Twelve artificial bone fragments (Sawbones, fourth-generation composite, item no: 3403 medium–left ) cut from the femur diaphysis model were used. The diameter and length of all sawbones were the same and were approximately 2 cm and 10 cm respectively. These fragments were bisected thoroughly in the coronal plane. Twelve CoCr Dall Miles (Med – Tıp, İzmir, Turkey) 2 mm diameter cable systems were used in two groups; the first group was the sleeve group (n = 6) and the second group was the sleeve + knot group (n = 6), focusing on 2 different cerclage wiring techniques. All cables were once looped around the saw bone with positioning the sleeve at the beaded end of the cable, and the tensioner knob was used to create tension to 90 lb/in2 (62,05 N/cm2).15 The sleeve was crimped with crimp tool. These procedures were the same for the two groups. In the sleeve + knot group, one more simple knot was performed with conventional surgery skill, and the two ends of cerclage wire were put into the tensioner knob to create tension again. Under tension, the two ends of the cable were cut using a cable cutter (Figure 1).Mechanical TestingMechanical testing was performed on a test system (Shimadzu Autograph AG-IS 5kN, Japan) with a 5 kN load-cell. The half of the saw bone fragments were mounted separately on two metallic half-cylinders forming a full cylinder. Each metal half-cylinder was fastened to a frame. The upper frame was fixated to the load cell of the system using a Schanz screw passing through all the holes in the frame and load cell. The lower frame was rigidly fixated to the base platform of the actuator. Both half metal cylinders mediatized and saw bone fragments were put on these cylinders to provide contact to each other (Figure 2). The cables were looped around the two cortical half shells, passed into the sleeve, tensioned and cut and manually pretensioned. Mechanical tension testing was performed at a velocity of 20 mm/min. Axial tension force was applied until failure occurred (Figure 3). The maximum force and displacement values were recorded at the failure point when the applied force suddenly dropped.Statistical AnalysisWe used the SPSS software (SPSS 26; SPSS, Chicago, USA) for statistical analysis. The ratio of maximum force and displacement was used to evaluate failure between the groups. In the analytical evaluation, the Mann-Whitney test was applied to evaluate the differences in the mean ranks of the two groups because the comparison of these two groups provided nonparametric conditions. The significance was defined as P < .05.
Results
All saw bone models were broken at the failure point. The amount of applied force leading to the initiation of the break of the saw bone models and the displacement of half shell pieces at the moment of break are shown in Table 1. There was a significant difference in the mean rank of applied force between the sleeve and the sleeve + knot groups (P < .001,025) (Table 2). The Sleeve + knot group needed much more force to catastrophically destroy the cable system than only sleeve group. But no statistical difference in the amount of displacement between the two groups was seen (P < .001,378) (Table 2). When we look at the force/displacement ratio, we found a significant difference between two groups: the sleeve + knot group was more durable than the sleeve only group (P < .001,01) (Table 3). No failure or loosening of cable in the crimp happened.
Discussion
In this study, we investigated that adding a simple knot technique to the standard procedure of the cable system made it more strength and stable biomechanically. In this setup, the saw bone fragments were transformed into half-shells by cutting in the coronal plane, and these pieces were mounted on metal bars to create separation under our control and measure the applied force and the amount of displacement. We defined the failure point as the applied axial loading exceeded the centripetal force of the wire around the saw bone. There may be seen loosening in the cerclage wire system after fixation of the fracture during the postoperative period.16-19 In this study, we think this problem could be reduced with a simple knot. It was seen that adding a knot to cable – domino fixation system increased the stability by approximately 40% more strength.
The most common places of failure on the wire are usually at stress risers (knots, twists, or kinking areas). The site of failure is usually located at potential stress risers.1 In our study, we added an extra knot on the sleeve part of the cable. Thus, we think this adding knot spread the stress area to a wider area, and this may help resist more strongly. We cut our wire just near the knot under the tension of the device. Therefore, we might lose our tension after cutting, but even though the sleeve + knot group is superior to only the sleeve group. This study showed us that a device that protects against tension after cutting wire is needed in this area.
We have made knots by hand first and then used a tensioner device to create tension on the knot. This is compatible with other studies. Harnroongroj et al..20 found that stiffness was related to the tightening of the construct; they found that to achieve the best stability, 200 N tension cerclage wire was needed. Wähnert et al..21 found that twisting under tension was beneficial for the installation of pre-tension.
Cutting the wire causes a significant loss of tension. Which part of wire is cut, also effects the tension, within the twist at the end of the wire.21 Rook et al..22 investigated first the twisted wire and then cutting below the breakage, in addition, without breakage, to reach the precise stable long-term tension, the strength of the wire is as important as fracture reduction and bone-to- bone contact. In this study, we proved that adding a basic knot method significantly enhances the load resistance of the cerclage and improves the cerclage fixation.
Limitations
The limitation of this study is that, in normal human physiology, the cable system always exposures cyclic loading by dynamic forces. Accordingly, the reason of failure is fatigue. But we perform force from the center to the periphery by increasing continuously until broken down. This shows us strength and durability. We defined failure of the system when it broke. In the literature, Lenz et al. 23 defined more than 3 mm axial displacement, Talbot et al. 24 defined displacement of 10 mm, or the point where applied force suddenly dropped by 10%. The point of only catastrophic failure of each system was determined. We also dealt only with centripetal force with axial loading.
Conclusion
This knot technique is very simple and fast. In our biomechanical study, we showed that only adding a knot to a cable system increases the durability and strength of the cable system by approximately 40%. These findings showed that adding a knot to the cable-domino system decreased the possibility of loosening and failing of fixation. There will be some advantages in clinical use considering the difficulties, especially in fixation of periprosthetic fractures, complex fractures using screws.
Declarations
Animal and Human Rights Statement
All procedures performed in this study were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. No animal or human studies were carried out by the authors for this article.
Data Availability
The datasets used and/or analyzed during the current study are not publicly available due to patient privacy reasons but are available from the corresponding author on reasonable request.
Conflict of Interest
None of the authors received any type of financial support that could be considered potential conflict of interest regarding the manuscript or its submission.
Funding
None.
References
- Dall DM. A biomechanical and clinical review: the Dall-Miles cable system. In: Sofue M, Endo N, eds. Treatment of Osteoarthritic Change in the Hip: Joint Preservation or Joint Replacement. Shinano, Japan: Springer; 2007:239-243. doi:10.1007/978-4-431-38200-3_24
- Craig NJA, Sivaji C, Maffulli N. Subtrochanteric fractures: a review of treatment options. Bull Hosp Jt Dis. 2001;60(1):35-46.
- Grimsrud C, Monzon RJ, Richman J, Ries MD. Cemented hip arthroplasty with a novel cerclage cable technique for unstable intertrochanteric hip fractures. J Arthroplasty. 2005;20(3):337-343. doi:10.1016/j.arth.2004.04.017
- Lin HH, Hung SH, Su YP, Chiu FY, Liu CL. Cerclage wiring in displaced associated anterior column and posterior hemi-transverse acetabular fractures. Injury. 2012;43(6):917-920. doi:10.1016/j.injury.2012.01.016
- Han KJ, Lee DH, Bang JY. Do cerclage cables delay the time to bone union in patients with an unstable humeral shaft fracture treated with intramedullary nails? Yonsei Med J. 2017;58(4):837-841. doi:10.3349/ymj.2017.58.4.837
- Müller T, Topp T, Kühne CA, Gebhart G, Ruchholtz S, Zettl R. The benefit of wire cerclage stabilisation of the medial hinge in intramedullary nailing for the treatment of subtrochanteric femoral fractures: a biomechanical study. Int Orthop. 2011;35(8):1237-1243. doi:10.1007/s00264-010-1204-4
- Angelini A, Battiato C. Combination of low-contact cerclage wiring and osteosynthesis in the treatment of femoral fractures. Eur J Orthop Surg Traumatol. 2016;26(4):397-406. doi:10.1007/s00590-016-1761-3
- Kulkarni SG, Babhulkar SS, Kulkarni SM, Kulkarni GS, Kulkarni MS, Patil R. Augmentation of intramedullary nailing in unstable intertrochanteric fractures using cerclage wire and lag screws: a comparative study. Injury. 2017;48(Suppl 2):S18-S22. doi:10.1016/s0020-1383(17)30489-8
- Angelini A, Battiato C. Past and present of the use of cerclage wires in orthopedics. Eur J Orthop Surg Traumatol. 2015;25(4):623-635. doi:10.1007/s00590-014-1520-2
- Liu A, O’Connor DO, Harris WH. Comparison of cerclage techniques using a hose clamp versus monofilament cerclage wire or cable. J Arthroplasty. 1997;12(7):772-776. doi:10.1016/s0883-5403(97)90007-3
- Dennis MG, Simon JA, Kummer FJ, Koval KJ, DiCesare PE. Fixation of periprosthetic femoral shaft fractures occurring at the tip of the stem: a biomechanical study of 5 techniques. J Arthroplasty. 2000;15(4):523-528. doi:10.1054/arth.2000.4339
- Dall DM, Miles AW. Re-attachment of the greater trochanter: the use of the trochanter cable-grip system. J Bone Joint Surg Br. 1983;65(1):55-59. doi:10.1302/0301-620x.65b1.6337168
- Kraus MJ, Nowakowski AM. Cerclage breakage in trochanteric revision—high risk for hip endoprosthesis? Acta Chir Orthop Traumatol Cech. 2017;84(2):138-140. doi:10.55095/achot2017/021
- Codesido P, Mejía A, Riego J, Ojeda-Thies C. Cerclage wiring through a mini-open approach to assist reduction of subtrochanteric fractures treated with cephalomedullary fixation: surgical technique. J Orthop Trauma. 2017;31(8):263-268. doi:10.1097/bot.0000000000000871
- Ménard J, Émard M, Canet F, Brailovski V, Petit Y, Laflamme GY. Initial tension loss in cerclage cables. J Arthroplasty. 2013;28:1509-1512. doi:10.1016/j.arth.2013.03.014
- Shin WC, Moon NH, Jang JH, Jeong JY, Suh KT. Technical note and surgical outcomes of percutaneous cable fixation in subtrochanteric fracture: a review of 51 consecutive cases over 4 years in two institutions. Injury. 2019;50(2):409-414. doi:10.1016/j.injury.2018.10.027
- Klinge SA, Vopat BG, Daniels AH, et al. Early catastrophic failure of trochanteric fixation with the Dall-Miles cable grip system. J Arthroplasty. 2014;29(6):1289-1291. doi:10.1016/j.arth.2014.01.001
- Kim JW, Park KC, Oh JK, Oh CW, Yoon YC, Chang HW. Percutaneous cerclage wiring followed by intramedullary nailing for subtrochanteric femoral fractures: a technical note with clinical results. Arch Orthop Trauma Surg. 2014;134(9):1227-1235. doi:10.1007/s00402-014-2023-7
- Meyer DC, Ramseier LE, Lajtai G, Nötzli H. A new method for cerclage wire fixation to maximal pre-tension with minimal elongation to failure. Clin Biomech (Bristol). 2003;18:975-980. doi:10.1016/s0268-0033(03)00181-5
- Harnroongroj T. Twist knot cerclage wire: the appropriate wire tension for knot construction and fracture stability. Clin Biomech (Bristol). 1998;13(6):449-451. doi:10.1016/s0268-0033(98)00005-9
- Wähnert D, Lenz M, Schlegel U, Perren S, Windolf M. Cerclage handling for improved fracture treatment: a biomechanical study on the twisting procedure. Acta Chir Orthop Traumatol Cech. 2011;78(3):208-214.
- Rooks RL, Tarvin GB, Pijanowski GJ, Daly WB. In vitro cerclage wiring analysis. Vet Surg. 1982;11:39-43. doi:10.1111/j.1532-950x.1982.tb00665.x
- Lenz M, Perren SM, Richards RG, et al. Biomechanical performance of different cable and wire cerclage configurations. Int Orthop. 2013;37(1):125-135. doi:10.1007/s00264-012-1702-7
- Talbot M, Zdero R, Schemitsch EH. Cyclic loading of periprosthetic fracture fixation constructs. J Trauma. 2008;64(5):1308-1312. doi:10.1097/ta.0b013e31811ea244
Tables
Table 1. Values of forces and displacements in sleeve and sleeve + knot groups
Table 2. Comparison of maximum forces and maximum displacement between two groups
*: Mann – Withney Test
Table 3. Comparison of ratio of force/displacement between two groups
*: Mann – Withney Test
Additional Information
Publisher’s Note
Bayrakol MP remains neutral with regard to jurisdictional and institutional claims.
Rights and Permissions
This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License (CC BY-NC 4.0). To view a copy of the license, visit https://creativecommons.org/licenses/by-nc/4.0/
About This Article
How to Cite This Article
Ertuğrul Şahin, Ahmet Karakaşlı, Bora Uzun, Hasan Havıtçıoğlu. Augmentation of cerclage wire strength with a basic knot technique: a biomechanical study. Ann Clin Anal Med 2021;12(2):191-194. doi:10.4328/ACAM.20456
Publication History
- Received:
- 26.12.2020
- Accepted:
- 11.01.2021
- Published Online:
- 15.01.2021
- Printed:
- 01.02.2021